Market access · ASEAN

ASEAN registration: one dossier format, six different reviews.

The ASEAN Medical Device Directive harmonised the CSDT structure and classification rules, not the approvals. This page explains what each authority in the region actually reviews, where test reports go in the dossier, and how to avoid a submission that is complete in format but thin in evidence.

AMDD / CSDT6 marketsLocal representative requiredTest evidence mapped to dossier sections
In short

A CSDT dossier is one document set reused across ASEAN, but each country registers separately through a local representative or licence holder. Test reports sit in the design verification and preclinical sections, and their completeness — not the country where they were produced — drives the review. An accredited laboratory report written in CSDT order saves weeks of correspondence, because the reviewer finds what the template asks for.

Where each market stands

Implementation status in 2026.

All ten member states signed the directive; maturity varies widely. The four below plus Vietnam and the Philippines carry most of the region's device volume.
ASEAN member state medical device regulatory status
CountryAuthorityAMDD statusPractical notes
SingaporeHSAFully implementedMost mature framework. Registration through PRISM, with immediate, expedited and full routes depending on class and prior reference approvals. Class A exempt.
MalaysiaMDAFully implementedLocal authorised representative required. Joined MDSAP in 2025. Conformity assessment route depends on class.
ThailandThai FDAFully implementedLicensed importer needed. CSDT accepted for Classes 2–4. Thai-language labelling and IFU.
IndonesiaMoH / BPOMFully implementedAKL registration through a local distributor. Indonesia reviews on its own terms even where CE or FDA approval exists.
VietnamMOH (DMEV)Partially implementedClass A/B at provincial level with short turnaround; C/D at MOH. Broad recognition of reference-agency approvals.
PhilippinesFDA (CDRRHR)TransitioningCSDT preferred under current circulars; notification for lower classes, registration for C and D.

Status compiled from HSA guidance and published AMDD/CSDT implementation summaries, November 2026. Always confirm with the authority or your local agent before filing.

Anatomy of a CSDT

Where the test reports actually go.

Manufacturers often send a folder of PDFs and call it a dossier. Reviewers work through the template in order, so the same content in the wrong section reads as missing.
CSDT sections and the evidence we supply
CSDT sectionWhat it containsOur input
Executive summaryDevice, intended use, classification, regulatory historyPlain-English summary of test scope and conclusions
Device descriptionMaterials, dimensions, components, accessories, variantsMaterial composition data; configuration list used in testing
Essential principles checklistApplicable safety and performance principles with evidence referencesClause-level references to safety and EMC reports
Risk analysisHazard identification, risk control, residual riskTest results feeding risk control verification
Design verification & validationPhysical, chemical, electrical, software and packaging evidenceFull test reports with raw data and method statements
Preclinical studiesBiological evaluation, animal studies, sterilizationISO 10993 reports, biological evaluation report, GLP studies
Clinical evaluationLiterature, clinical data or equivalence justificationGap memo on what clinical evidence is still needed
Labelling & IFULabels, instructions, translationsLabel content verification where symbols are claimed
QMS & manufacturingISO 13485 or equivalent, site information, process flowSterilization and packaging process validation

Sequencing

Testing first, filing second — but plan them together.

The slowest part of an ASEAN submission is usually not the authority's review clock, it is the gap between filing and the arrival of a report nobody scheduled. Build the test calendar backwards from your target filing date.

  • Week 0: classification per country, since a device can land in different classes in different markets.
  • Weeks 1–2: test plan and gap review against the CSDT template, including reports you already hold.
  • Weeks 3–14: biological evaluation, safety/EMC, sterilization and packaging work running in parallel where the device allows.
  • Filing: dossier assembled with our index table; local agent submits; deficiency responses come back to a single technical contact.
What we do not do: hold a local representative licence or act as your agent in any ASEAN market. Registration filing and authority liaison sit with your appointed local agent or our regulatory partners; we supply the technical evidence they file.
Standard Group medical device laboratory where CSDT test evidence is generated
SINGLE QUALITY SYSTEM · MULTI-MARKET EVIDENCE

FAQ

ASEAN registration questions.

Is one ASEAN registration valid across all ten countries?

No. The directive harmonises dossier format and classification, but each member state issues its own registration and requires its own local representative or licence holder. One CSDT can be reused; the approval is country by country.

Which CSDT sections do you write?

The technical ones: design verification and validation, preclinical summaries, biological evaluation, sterilization and packaging validation, plus a clause-level index into the essential principles checklist. Device description and clinical sections are usually owned by the manufacturer; we flag gaps in both.

How long does registration take?

Vietnam Class A/B can clear in one to two weeks at provincial level; Singapore's immediate route is quick where a reference approval exists. Malaysia, Thailand, Indonesia and the Philippines typically take months, and the clock starts only when the file is complete.

Do we need ISO 13485?

In practice yes. HSA accepts ISO 13485, MDSAP, US FDA QSR conformity or Japan MHLW Ordinance 169; MDA and Thai FDA expect equivalent evidence. Test data cannot compensate for a missing quality system.

Can we start from an existing CE or FDA file?

Often yes — the underlying test data is reusable if the standards and versions match what the ASEAN authority recognises. What changes is structure, language and the essential principles mapping. We review the existing file and tell you what can move across unchanged.

Next step

Send your target markets and your current file.

We return a gap list against each authority's expectations, plus the tests still needed and their dates.

Request a gap review standard@sitande.com Replies within one working day · English