Dossier structure

The CSDT dossier: 12 items Southeast Asian reviewers open first.

The ASEAN Common Submission Dossier Template is a fixed structure. Reviewers work through it in order, which means content in the wrong section reads as missing content. Here is what goes where, and where files most often stall.

Key takeaways

  • The CSDT format is harmonised; the approval is not — each country registers separately.
  • Test reports belong in the design verification and preclinical sections, referenced from the essential principles checklist.
  • Most deficiencies come from the clinical evaluation, the QMS evidence and the labelling section — not from failed bench tests.
  • Assembling the file in reverse order (evidence first, summary last) prevents the "complies, no reference" problem.

The twelve sections

A CSDT submission is built from a common set of documents. The table below is the working order used by most ASEAN authorities, and the column on the right is what we typically supply versus what the manufacturer or local agent owns.

CSDT sections and ownership
#SectionSupplied by
1Executive summaryTogether — we draft the technical summary, you confirm intended use
2Device descriptionManufacturer; we provide material composition data
3Essential principles checklistTogether — clause-level references to our reports
4Risk analysis and controlManufacturer; our results verify risk controls
5Design verification — physical and chemicalUs: test reports with protocols and raw data
6Design verification — electrical and softwareUs for safety and EMC; manufacturer for software lifecycle
7Preclinical studiesUs: biological evaluation, animal studies, sterilization
8Clinical evaluationManufacturer or clinical consultant; we flag evidence gaps
9Labelling and IFUManufacturer plus translation; we verify claimed symbols
10QMS evidenceManufacturer: ISO 13485, MDSAP or equivalent
11Manufacturing and sterilization sitesManufacturer; we provide process validation
12Declaration of conformityManufacturer's authorised signatory

Where dossiers come back incomplete

In our experience the same three areas generate most deficiency letters. None of them is usually a testing problem.

1. The essential principles checklist has no evidence trail

Each principle needs a reference to a specific report, clause and page. "Complies with ISO 10993" is not a reference. A reviewer who cannot trace the principle to data will ask for the data — and the clock restarts even if the data exists.

2. The clinical evaluation is thin for a well-tested device

Manufacturers spend heavily on bench testing and then submit three literature papers with no search strategy, no appraisal criteria and no equivalence argument. For Class C and D devices this is the most common cause of a major deficiency.

3. Sterilization and packaging evidence does not match the marketed configuration

A validation report covering one pouch size while the device ships in three is an easy objection. The validation has to name the product family, the worst-case configuration and the load pattern, or it will be treated as incomplete.

Assembling the file in the right order

Write the evidence documents first and the executive summary last. In practice that means four passes:

  1. Index the evidence. List every report with its standard, edition, report number and the principles it supports.
  2. Draft the checklist from the index. Every row gets a report reference before the sentence is written.
  3. Write the preclinical summary. Ten to fifteen pages that a reviewer can read without opening the underlying reports.
  4. Write the executive summary. Device, intended use, class per market, evidence map, open points. One page if possible.

This order feels slower for the first two days and reliably saves weeks afterwards, because the gaps surface while there is still time to schedule a test slot.

A one-page pre-filing check

  • Device classification stated per country, not once for the region.
  • Every essential principle row has a report, clause and page reference.
  • Biocompatibility plan states contact category, duration and extraction justification.
  • Sterilization validation names the product family, worst case and load pattern.
  • Packaging validation includes ageing and distribution, not only seal strength.
  • Software version in the file matches the version tested.
  • Labels and IFU translated for each market, with symbols consistent with the standards claimed.
  • QMS certificate valid at the intended filing date, not at the date of printing.
  • Clinical evaluation includes search strategy, appraisal and equivalence argument.
  • All reports are signed, dated and traceable to the laboratory's accreditation scope.

If two or more of these are weak, the file is not ready — regardless of how good the test data is. The related pages below cover the testing side of each item in more detail.


About this note. Written by the Standard Group Southeast Asia desk from project work on ASEAN submissions. Regulatory requirements change; confirm current guidance with the relevant authority or your local agent before filing. Last reviewed 18 November 2026.