Biological evaluation plan
Contact category, materials, endpoints selected and explicitly excluded, with justification and the extraction rationale.
Biological evaluation · ISO 10993
A biological evaluation plan, the in vitro and in vivo tests your contact category requires, and chemical characterization that explains what the device releases — documented to ISO 10993-1:2025 and structured for CSDT, MDR and 510(k) review.
Biocompatibility testing is a risk-based evaluation, not a fixed menu. ISO 10993-1:2025 asks for chemical and physical characterization first, then in vitro and in vivo tests selected by contact category and duration. For most devices that means cytotoxicity, irritation or sensitization, plus ISO 10993-18 chemical characterization — with implantation or systemic toxicity added for implants and long-term contact.
How the plan is built
| Contact category | Duration | Core endpoints usually required | Typically added |
|---|---|---|---|
| Surface device skin, mucosa, breached surface | Limited (≤24 h) Prolonged (24 h–30 d) | Cytotoxicity; irritation; sensitization | Chemical characterization for mucosa or breached-surface contact; clinical use justification |
| External communicating indirect blood path, tissue-fluid, blood path | Limited / prolonged / long-term | Cytotoxicity, irritation, sensitization, plus hemocompatibility for blood contact | Systemic toxicity, implantation, chemical characterization, EO residuals where applicable |
| Implant device tissue, bone, blood | Prolonged / long-term | Cytotoxicity, sensitization, implantation with histology, chemical characterization | Systemic toxicity, genotoxicity, carcinogenicity justification, degradation and metal-ion release studies |
| Resorbable / degradable | All | Degradation product characterization, local effects after implantation | Absorption, metabolic pathways, long-term implant endpoints |
The matrix is a starting point. Final scope depends on materials, surface area, manufacturing residues, sterilization and the intended clinical exposure.
In vitro & in vivo
Cell culture and animal work sit in the same quality system as the analytical laboratory, so sample preparation, extraction and results stay traceable across the whole evaluation.
ISO 10993-18
Chemical characterization identifies what a device is made of and what it releases under clinical conditions. Done properly, it can remove whole in vivo endpoints; done superficially, it produces a report that reviewers reject and a re-test bill.
Deliverables
Contact category, materials, endpoints selected and explicitly excluded, with justification and the extraction rationale.
Signed, accredited reports per method with raw data, instrument traces, photomicrographs and deviations noted.
A consolidated conclusion tying chemical data, in vitro and in vivo results into one biological safety statement.
Exposure-based assessment of identified substances with tolerable intake comparison and open uncertainties stated.
Reports indexed to the dossier sections your agent or notified body will open, in English, with an index table.
Short letter listing what is still missing for your target market — clinical evidence, labelling, sterilization, software.
FAQ
It often reduces it, but it does not automatically replace it. ISO 10993-1:2025 expects characterization first, with in vivo tests added where the chemical data cannot establish biological safety — novel materials, long-term implants, uncertain degradation. Cytotoxicity is still expected in almost every case.
Per ISO 10993-12, chosen to be at least as severe as clinical use: 37 °C / 24 h for short-term surface contact, 37 °C / 72 h for prolonged contact, and elevated conditions such as 50 °C / 72 h or 70 °C / 24 h where materials allow. Conditions are justified in the plan and stated in the report.
Only where biological equivalence holds: same materials, same manufacturing process, same contact category and duration, same sterilization. A change of resin supplier, mould release agent, colourant or sterilization method usually requires at least partial re-testing.
It scales with exposed surface area, so a dental implant and a surgical drape need very different quantities. After the plan is agreed we issue a sample request for finished devices in final packaging.
Yes. Reports are written in CSDT structure with an executive summary of preclinical studies, which is the section HSA, MDA and Thai FDA reviewers open first. See our ASEAN registration page.
Next step
You get a plan that states which endpoints apply, which do not, and why — with dates and price.